RT Journal Article T1 Intraoperative computed tomography imaging for dose calculation in intraoperative electron radiation therapy: Initial clinical observations A1 García Vázquez, Verónica A1 Calvo, Felipe A A1 Ledesma Carbayo, María Jesús A1 Vicente Solé, Claudio A1 Calvo Haro, José Antonio A1 Desco Menéndez, Manuel A1 Pascau González-Garzón, Javier AB In intraoperative electron radiation therapy (IOERT) the energy of the electron beam is selected under the conventional assumption of water-equivalent tissues at the applicator end. However, the treatment field can deviate from the theoretic flat irradiation surface, thus altering dose profiles. This patient-based study explored the feasibility of acquiring intraoperative computed tomography (CT) studies for calculating three-dimensional dose distributions with two factors not included in the conventional assumption, namely the air gap from the applicator end to the irradiation surface and tissue heterogeneity. In addition, dose distributions under the conventional assumption and from preoperative CT studies (both also updated with intraoperative data) were calculated to explore whether there are other alternatives to intraoperative CT studies that can provide similar dose distributions. The IOERT protocol was modified to incorporate the acquisition of intraoperative CT studies before radiation delivery in six patients. PB PLOS SN 1932-6203 YR 2020 FD 2020-01-10 LK https://hdl.handle.net/10016/31915 UL https://hdl.handle.net/10016/31915 LA eng NO This study was supported by Ministerio de Ciencia, Innovación y Universidades (http://www.ciencia.gob.es) [grant number TEC2013–48251-C2 to JP, VG-V and MJL-C], co-funded by European Regional Development Fund (ERDF), “A way of making Europe” (https://ec.europa.eu/regional_policy/en/funding/erdf); by Ministerio de Ciencia, Innovación y Universidades (http://www.ciencia.gob.es), Instituto de Salud Carlos III (https://www.isciii.es) [grant numbers DTS14/00192 to JP, VG-V and FAC; PI15/02121 to FAC and JC-H; PI18/01625 to JP], co-funded by European Regional Development Fund (ERDF), “A way of making Europe” (https://ec.europa.eu/regional_policy/en/funding/erdf); and by Comunidad de Madrid (http://www.comunidad.madrid) [grant number TOPUS-CM S2013/MIT3024 to JP], co-funded by European Structural and Investment Fund (https://ec.europa.eu/info/funding-tenders/funding-opportunities/funding-programmes/overview-funding-programmes_en). The CNIC is supported by the Ministerio de Ciencia, Innovación y Universidades (http://www.ciencia.gob.es) and the Pro CNIC Foundation (https://www.fundacionprocnic.es) [to MD], and is a Severo Ochoa Center of Excellence (SEV-2015-0505). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript. DS e-Archivo RD 27 jul. 2024