Publication:
Real-Time Digital Timing in Positron Emission Tomography

Loading...
Thumbnail Image
Identifiers
Publication date
2008-12-08
Defense date
Advisors
Tutors
Journal Title
Journal ISSN
Volume Title
Publisher
IEEE
Impact
Google Scholar
Export
Research Projects
Organizational Units
Journal Issue
Abstract
Positron emission tomography (PET) requires accurate timing of scintillation events to properly discriminate between coincident and noncoincident pairs. The traditional solution to timing is based on custom application specific integrated circuits (ASIC) designs, whose cost may not be justified in the design of experimental small animal PET scanners. The new generation of PET scanners introduces the idea of continuous sampling of the detected scintillation pulse, replacing event-triggered acquisition front-ends. This approach enables new options to the timing procedure based on digital processing of the sampled pulse signal. This work proposes a time stamping algorithm based on the optically matched filter and compares the potential performance benefits of this approach versus other FIR-based timing algorithms, some of which have been already implemented by different authors. Results show that the coincidence timing resolution may be as low as 1.5 ns without the need of expensive high-speed converters when the proper signal processing is applied
Description
Keywords
Data acquisition, Medical signal processing, Positron Emission Tomography (PET)
Bibliographic citation
IEEE Transactions on Nuclear Science, 2008, vol. 55, n. 5, p. 2531 - 2540