Archivo Abierto Institucional de la Universidad Carlos III de Madrid >
Investigación >
Departamentos >
Departamento de Bioingeniería e Ingeniería Aeroespacial: Bioingeniería >
DBIAB - Book Chapters >
Please use this identifier to cite or link to this item:
http://hdl.handle.net/10016/12219
|
Google™ Scholar.
Others By:
Sisniega, Alejandro
- Vaquero, Juan José
- Lage, Eduardo
- Carlos, Álvaro de
- Villena, José L.
- Abella, Mónica
- Vidal, Irina
- Tapias, Gustavo
- Antoranz, José Carlos
- Desco, Manuel
|
| Title: | Comparative study of two flat-panel x-ray detectors applied to small-animal imaging cone-beam micro-CT |
| Author(s): | Sisniega, Alejandro Vaquero, Juan José Lage, Eduardo Carlos, Álvaro de Villena, José L. Abella, Mónica Vidal, Irina Tapias, Gustavo Antoranz, José Carlos Desco, Manuel |
| Publisher: | IEEE |
| Issued date: | Oct-2008 |
| Citation: | 2008 IEEE Nuclear Science Symposium Conference Record, Oct. 2008, p. 3836-3840 |
| URI: | http://hdl.handle.net/10016/12219 |
| ISBN: | 978-1-4244-2714-7 |
| ISSN: | 1082-3654 |
| DOI: | http://dx.doi.org/10.1109/NSSMIC.2008.4774118 |
| Description: | Proceeding of: 2008 IEEE Nuclear Science Symposium Conference Record (NSS '08), Dresden, Germany, 19-25 Oct. 2008 |
| Abstract: | This work compares two different X-ray flat-panel detectors for its use in high-speed, cone-beam CT applied to small-animal imaging. The main differences between these two devices are the scintillators and the achievable frame rate. Both devices have been tested in terms of system linearity, sensitivity, resolution, stability and noise properties, taking into account the different timing schemes for each one of them and the mandatory corrections on the raw data. Tomographic scans have been carried out using both detectors to evaluate its final performance as well as the delivered dose needed to achieve similar quality scans. An experimental cone-beam CT test-bench has been designed and implemented to perform the different measurements. It uses a micro-focus X-ray source and a rotating stage where the samples are placed. A modified FDK algorithm has been used to reconstruct the acquired data. Both detectors show similar results for pixel linearity and stability measurements, and their noise levels are comparable. The resolution and sensitivity features are better for the direct grown scintillator detector (9 Ipmm vs. 6 Ipmm, and -4 times more sensitive for similar delivered dose). Since tomographic reconstructed images for the higher frame-rate detector show acceptable quality, it can be used to implement a faster system for high-speed acquisition techniques like, for example, dynamic imaging or gated protocols. |
| Sponsor: | This work is partially funded by the CD-TEAM Project, CENIT Program, Spanish Ministerio de Industria, and with grants from the Ministerio de Educacion y Ciencia, projects TEC2007-64731 and TEC2008-06715-C02Ol. |
| Publisher version: | http://dx.doi.org/10.1109/NSSMIC.2008.4774118 |
| Rights: | © IEEE |
| Appears in Collections: | DBIAB - Proceedings DBIAB - Book Chapters
|
Items in E-Archivo are protected by copyright, with all rights reserved, unless otherwise indicated.
|